<Error>
<ErrorNumber>22017</ErrorNumber>
<ErrorFamily>Autoloader 2</ErrorFamily>
<ErrorMessage>
ALE2: System air pressure too low!
</ErrorMessage>
<AppliesTo>
<UnnumberedList>
<ListItem><Text>ThermoFlex</Text></ListItem>
</UnnumberedList>
</AppliesTo>
<ServiceInformation>
<DescriptionText><text>
Check sensing line connection to TEE board, a loud hissing would signify a
loose connection.  The TEE board may be faulty, check for line pressure before
replacing board.  Use of the ale command will provide the current pressure reading.
</text></DescriptionText>
</ServiceInformation>
<OperatorAction>
<DescriptionText><text>
Ensure pressurized air is connected.
Check Pneumatic regulator in machine to verify pressure.
If all is in order, call Creo service.
</text></DescriptionText>
</OperatorAction>
<Cause>
The air pressure sensed at the TEE board has dropped below the threshold.
The air pressure threshold is also modified by the air supply offset parameter
in the al nvs sub-system.
One other cause is that the table float or vacuum system is causing enough
pressure drop to the system that the threshold is not met.
</Cause>
<SeeAlso>
<UnnumberedList>
<Link><LinkAddress>../commands/ale.xml</LinkAddress><LinkTitle>Command: ale</LinkTitle></Link>
<Link><LinkAddress>../nvs/sys_cfg_spt.xml</LinkAddress><LinkTitle>NVS: sys spt</LinkTitle></Link>
<Link><LinkAddress>../nvs/al_cal_aso.xml</LinkAddress><LinkTitle>NVS: al aso</LinkTitle></Link>
</UnnumberedList>
</SeeAlso>
<OtherLinks>
<Link><LinkAddress>..\..\ShowAllErrors.htm</LinkAddress><LinkTitle>Firmware 0 Error Index</LinkTitle></Link>
<Link><LinkAddress>..\..\ShowAllCommands.htm</LinkAddress><LinkTitle>Firmware 0 Command Index</LinkTitle></Link>
<Link><LinkAddress>..\..\ShowAllNVS.htm</LinkAddress><LinkTitle>Firmware 0 NVS Index</LinkTitle></Link>
</OtherLinks>
<Updated>
Updated: Wed Jul 16 19:33:51 2008 PST
</Updated>
<Links>
</Links>
</Error>